wp--preset--color--luminous-vivid-amber) !important;}.has-light-green-cyan-border-color{border-color: var(--wp--preset--color--light-green-cyan) !important;}.has-vivid-green-cyan-border-color{border-color: var(--wp--preset--color--vivid-green-cyan) !important;}.has-pale-cyan-blue-border-color{border-color: var(--wp--preset--color--pale-cyan-blue) !important;}.has-vivid-cyan-blue-border-color{border-color: var(--wp--preset--color--vivid-cyan-blue) !important;}.has-vivid-purple-border-color{border-color: var(--wp--preset--color--vivid-purple) !important;}.has-vivid-cyan-blue-to-vivid-purple-gradient-background{background: var(--wp--preset--gradient--vivid-cyan-blue-to-vivid-purple) !important;}.has-light-green-cyan-to-vivid-green-cyan-gradient-background{background: var(--wp--preset--gradient--light-green-cyan-to-vivid-green-cyan) !important;}.has-luminous-vivid-amber-to-luminous-vivid-orange-gradient-background{background: var(--wp--preset--gradient--luminous-vivid-amber-to-luminous-vivid-orange) !important;}.has-luminous-vivid-orange-to-vivid-red-gradient-background{background: var(--wp--preset--gradient--luminous-vivid-orange-to-vivid-red) !important;}.has-very-light-gray-to-cyan-bluish-gray-gradient-background{background: var(--wp--preset--gradient--very-light-gray-to-cyan-bluish-gray) !important;}.has-cool-to-warm-spectrum-gradient-background{background: var(--wp--preset--gradient--cool-to-warm-spectrum) !important;}.has-blush-light-purple-gradient-background{background: var(--wp--preset--gradient--blush-light-purple) !important;}.has-blush-bordeaux-gradient-background{background: var(--wp--preset--gradient--blush-bordeaux) !important;}.has-luminous-dusk-gradient-background{background: var(--wp--preset--gradient--luminous-dusk) !important;}.has-pale-ocean-gradient-background{background: var(--wp--preset--gradient--pale-ocean) !important;}.has-electric-grass-gradient-background{background: var(--wp--preset--gradient--electric-grass) !important;}.has-midnight-gradient-background{background: var(--wp--preset--gradient--midnight) !important;}.has-small-font-size{font-size: var(--wp--preset--font-size--small) !important;}.has-medium-font-size{font-size: var(--wp--preset--font-size--medium) !important;}.has-large-font-size{font-size: var(--wp--preset--font-size--large) !important;}.has-x-large-font-size{font-size: var(--wp--preset--font-size--x-large) !important;} :where(.wp-block-post-template.is-layout-flex){gap: 1.25em;}:where(.wp-block-post-template.is-layout-grid){gap: 1.25em;} :where(.wp-block-columns.is-layout-flex){gap: 2em;}:where(.wp-block-columns.is-layout-grid){gap: 2em;} :root :where(.wp-block-pullquote){font-size: 1.5em;line-height: 1.6;}

日立CR21G离心机开机出现 er13”错误

日立CR21G离心机的常见故障包括:

(1) 转动系统故障

(2) 真空系统密封不严。由于只有超速离心机 (55000~ 150000r/min) 才需配备真空装置, 所以该类故障仅存在于超速离心机

(3) 制冷系统故障。其中以转动系统故障最为常见。现在我们以日立 CR21G 落地式冷冻离心机为例, 予以介绍。

 

故障现象: 开机出现 er13”错误提示, 主机无法启动。

故障分析与检修: 该离心机出现 rr13”报警,原因可能在于以下两个方面。

(1)主机转头运行时,电路主控制板检测不到MPG信号。

( 2)转头运行时,重复3次检测到 MPG 信号超过 700min- 1 的频率改变。

打开离心腔,移去3个M4×12固定螺钉,取出传感器盖板,暴露传感器。用手转动驱动主轴,万用表检测电路中 18MPG 的电压变化,如图1所示。

 

若未发现周期性电压变化,再确定P211接口是否接触良好。离心机正常工作情况下, 可在18MPC检测到工作波形,如图2所示。

 

MPG传感器,即霍尔效应元件,此处用于监测驱动转轴转速。2个线圈分别被固定在转轴底端对称部位,如图3。转轴每转动1圈, MPG可检测到2个方形脉冲。

 

检查发现P211接口接触正常,重新开机,rr13”报警依旧无法消除,则更换MPG元件,  更换后,将离心机调水平,开机,启动,离心机工作正常。

相关新闻

联系我们

phone-logo 400-1181176

返回顶部
服务热线:400-1181176